Balloon Inflation Pump Special Pressure Gauge B21.02C
thumbnail 1
thumbnail 2
thumbnail 3

Type Ⅲ-2 Balloon Inflation Pump Digital Pressure Gauge

Battery-powered digital balloon inflation gauge with peak-hold records and alarms.

Model: Type Ⅲ-2|Size: 40 mm [1½"]

The GARHOPE Type Ⅲ-2 balloon inflation pump digital pressure gauge is a fully digital medical pressure-monitoring instrument, dial size 40 [1½"], for one-time balloon dilation pressure pumps. It provides a 3-digit display with atm/psi unit switching, backlight, "VAC" negative-pressure indication, timing, peak-hold records and audible-visual over-pressure alarms, housed in a medical-grade ABS case with medical EP internal cleanliness and ≤1×10⁻⁹ Pa·m³/s gas tightness.

Key highlights:

  • 3-digit display with atm/psi unit switching; backlight; "VAC" mark and value under negative pressure; black-background white-text display mode.
  • Timing function — start on button press, stop when pressure falls to 0, auto-return to zero.
  • Peak-record function — stores the maximum inflation pressure and the holding time; up to 3 peak records.
  • Alarm function — over-limit alarm from 24 atm (orange flashing digits, backlight, buzzer ≥ 55 dB(A)); above 31 atm shows fault code E-H; alarm auto-clears below the setpoint.
  • Physical power-off via the on/off key; screen dims after 3 min and sleeps after 15 min with no pressure change.

Specifications

ItemSpecification
ModelType Ⅲ-2 (series Y40Z)
Dial size40 mm [1½"]
Range-1 ~ 0 ~ +30 ATM
Accuracy1% F.S (±0.3 ATM)
Overload pressure1.5 F.S (45 ATM)
Measuring mediumWater fluid
Sampling rate≥ 20 Hz; display update ≥ 5 Hz (≤ 200 ms between updates); no jumping to final value
Power2 × 3 V coin cells (CR2032) / external 5 V; ≥ 48 h battery operation; cell shelf life ≥ 5 years (stored, unpowered)
CaseMedical-grade white or black ABS
Measuring elementPressure sensor
ConnectionIntegrated with case (nylon & plastic)
Internal cleanlinessMedical EP grade
Gas tightness≤ 1×10⁻⁹ Pa·m³/s
Water resistanceTarget IP67 (fallback IP55 pending optimisation)
Operating temperature+10 °C to +55 °C (EO sterilisation at 55 °C)
Storage temperature-20 °C to +70 °C
Storage humidity≤ 95%
Data storage3 records

Applications

Balloon inflation pump pressure monitoring

Interventional procedures requiring peak-pressure and holding-time records

Medical device OEM assembly

Ordering & Customization

Supports batch supply and customized configurations. Medical OEM supported (72 h samples, 15-day mass production).